Numéro |
Rev. Phys. Appl. (Paris)
Volume 12, Numéro 2, février 1977
|
|
---|---|---|
Page(s) | 313 - 319 | |
DOI | https://doi.org/10.1051/rphysap:01977001202031300 |
DOI: 10.1051/rphysap:01977001202031300
Finite solid angle correction factors and efficiencies for cadmium telluride detectors
A.K. Chaubey et H.V. GuptaDepartment of Physics, Aligarh Muslim University Aligarh 202001, India
Abstract
Finite solid angle correction factors and absolute detection efficiencies of the CdTe detectors of right circular cylindrical geometry for point sources placed on the axis of the detector, has been calculated for γ-ray energies from 100 keV to 10 MeV, taking into account various source to detector distances, ranging from 1 to 10 cm. CdTe detectors of several area between 0.12 and 2.5 cm2 of various thicknesses were considered.
Résumé
On calcule les facteurs de correction d'angle solide et les efficacités de détection de compteurs CdTe de forme cylindrique droite pour des sources de rayonnements γ ponctuelles d'énergies comprises entre 100 keV et 10 MeV placées le long de l'axe du cylindre. Différentes distances source-détecteur, allant de 1 à 10 cm ont été considérées ainsi que des diodes de différentes surfaces, comprises entre 0,12 et 2,5 cm2 et de différentes épaisseurs.
2930K - X and gamma ray spectroscopy.
2940P - Semiconductor detectors.
Key words
gamma ray spectrometers -- semiconductor counters -- absolute detection efficiencies -- CdTe detectors -- right circular cylindrical geometry -- point sources -- 100 keV to 10 MeV -- gamma rays -- finite solid angle correction factors