Numéro
Rev. Phys. Appl. (Paris)
Volume 1, Numéro 1, mars 1966
Page(s) 74 - 76
DOI https://doi.org/10.1051/rphysap:019660010107400
Rev. Phys. Appl. (Paris) 1, 74-76 (1966)
DOI: 10.1051/rphysap:019660010107400

Détecteur de rapport pour un récepteur monocanal application à un monochromateur à simple faisceau

Amouyeli, Blet et Riollet

Laboratoire de Physique du Solide, Compagnie Française Thomson-Houston


Abstract
In order to measure the variation of some quantity as a function of one parameter with respect to a reference, one ordinary make use of a double channel instrument recording the ratio of the two values. We present here a new system, which allow us to make use of a single channel apparatus owing to an alternating exploration and an electromechanical memory triggered to the reference value. The special application described below corresponds to the modification of a single beam optical monochromator in a automatic spectrophotometer.


Résumé
Lorsque l'on veut mesurer les variations d'une grandeur par rapport à une référence en fonction d'un paramètre, on utilise de préférence un appareil à deux canaux dont on enregistre le rapport des indications. Le dispositif que nous présentons ici permet d'utiliser un appareil monocanal grâce à une exploration alternée et une mémoire électromécanique asservie à la grandeur de référence. L'application particulière décrite correspond à la transformation d'un monochromateur optique simple en spectrophotomètre automatique.

PACS
0707H - Display and recording equipment, oscilloscopes, TV cameras, etc..

Key words
Monochromators -- One channel system -- Data loggers -- Measurement method -- Transmittance -- Color filter -- Single-channel receiver