Dilatomètre linéaire capacitif à usages multiples

E. Du Trémolet de Lacheisserie

Laboratoire de Magnétisme, Université Paris VI, France


Abstract
This dilatometer allows changes of length between 0.5 Å and 100 μm to be measured in the temperature range 4.2-1 000 K ; the absolute accuracy is 1 %. Some examples of measurements are shown, concerning magnetostriction, piezoelectricity and thermal expansion. Since the sample can be stressed during the experiment, or heat-treated in situ in an applied field, it is also possible to study the directional order under good experimental conditions (english translation available).


Résumé
Ce dilatomètre permet de mesurer avec précision (1 %) tout allongement compris entre 0,5 Å et 100 μ que peut subir un échantillon solide entre 4,2 et 1 000 K. Quelques exemples d'utilisation de ce dilatomètre sont présentés : mesures de magnétostriction, de piézoélectricité, de dilatation thermique. La possibilité d'appliquer aux échantillons une contrainte uniaxiale en cours d'expérience ou un traitement thermique sous champ in situ, permet d'aborder les problèmes d'ordre directionnel dans d'excellentes conditions expérimentales.

PACS
0630C - Spatial variables measurement.
0750 - Electrical instruments and techniques.
6570 - Thermal expansion and thermomechanical effects.
7580 - Magnetomechanical and magnetoelectric effects, magnetostriction.
7760 - Piezoelectricity and electrostriction.

Key words
extensometers -- magnetostriction -- piezoelectricity -- thermal expansion -- 4.2 to 1000K -- multipurpose linear capacitance dilatometer -- absolute accuracy