Comparaison précise de courants périodiques à l'aide de thermistances à chauffage indirect
J.C. Bernengo1, 2, M. Hanss1, 2 et P. Rottembourg1, 21 Laboratoire de Biophysique du Muséum National d'Histoire Naturelle, Paris
2 Laboratoire de Chimie Macromoléculaire de la Faculté des Sciences de Lyon
Abstract
A very low frequency thermal transfer unit is described. This device is based on the use of two indirectly-heated thermistors, in two adjacent arms of a Wheatstone bridge. A periodic current (of any shape) can thus be compared to a known current (periodic or D.C.) with a relative sensitivity which may reach, under the best conditions, a few parts in 10-5 between 0,1 and 500 Hz. The bridge is still usable at much higher frequencies.
Résumé
On décrit un pont équipé de deux thermistances à chauffage indirect qui permet la comparaison précise d'un courant périodique (de forme quelconque) à un courant connu (périodique ou non). Dans les conditions les plus favorables, la sensibilité relative peut atteindre 10-5, entre 0,1 et 500 Hz. Ce dispositif est encore utilisable à des fréquences beaucoup plus élevées.