Numéro |
Rev. Phys. Appl. (Paris)
Volume 18, Numéro 6, juin 1983
|
|
---|---|---|
Page(s) | 355 - 358 | |
DOI | https://doi.org/10.1051/rphysap:01983001806035500 |
DOI: 10.1051/rphysap:01983001806035500
Stabilisation de la température d'un cryostat entre 1,6 K et 270 K pour spectrométrie dans l'infrarouge lointain
B. Wyncke, P. Strimer, M. El Shérif, F. Bréhat et C. JacquotLaboratoire Infrarouge Lointain, Université de Nancy I, B.P. 239, 54506 Vandoeuvre-les-Nancy Cedex, France
Abstract
In order to evaluate the variation of the lattice mode frequencies of single crystals with temperature a conventional infrared grating spectrometer has been modified. A regulated temperature cryostat is adapted to the apparatus ; a transmission and a reflexion cell are described. All over the 1.6 to 270 K temperature range the performance of the spectrometer was preserved. The Golay cell was replaced by a cooled gallium doped germanium bolometer and thus the sensibility of the spectrometer is increased by 10.
Résumé
Afin de pouvoir suivre d'une manière continue l'évolution des fréquences des modes de réseau de monocristaux en fonction de la température nous avons été amenés à modifier un spectromètre pour l'infrarouge lointain. Tout en conservant les performances de l'appareil nous avons adapté un cryostat à température variable (1,6-270 K). Une cellule de mesure pour l'étude de la transmission et une cellule pour l'étude de la réflexion sont décrites. La sensibilité du récepteur a été améliorée en remplaçant le détecteur de Golay par un bolomètre au germanium dopé au gallium.
0720M - Cryogenics.
0765G - IR spectroscopy and spectrometers.
Key words
cryostats -- interference spectrometers -- spectrometer components and accessories -- reflectivity measurements -- far infrared -- lattice mode frequencies -- infrared grating spectrometer -- regulated temperature cryostat -- transmission -- 1.6 to 270K -- Golay cell