Issue
Rev. Phys. Appl. (Paris)
Volume 1, Number 2, juin 1966
Page(s) 113 - 116
DOI https://doi.org/10.1051/rphysap:0196600102011301
Rev. Phys. Appl. (Paris) 1, 113-116 (1966)
DOI: 10.1051/rphysap:0196600102011301

Interféromètre de Michelson pour mesures diélectriques en ondes millimétriques

A. Gerschel et C. Brot

Laboratoire de Chimie Physique, Faculté des Sciences, Orsay


Abstract
A Michelson-type interferometer is described suitable for the measurement of dielectric constants and losses at 2 mm wavelength. The beam-dividing device is an air gap of adjustable thickness situated between two teflon prisms, whose faces undergo Brewsterian incidence. The planar sample, in either liquid or solid phase, lies at the bottom of a metallic cell, at the end of one of the arms of the interferometer. The accuracy of measurements varies inversely with the dielectric losses of the material, being less than 3 % for ε' and 6 % for ε", in the case of average losses. The dielectric constant of solid tert.-butyl chloride at - 35°C is found to be ε' = 2.7 and ε" = 1.7.


Résumé
On décrit un interféromètre du type Michelson pour la mesure de la permittivité et des pertes diélectriques à la longueur d'onde de 2 mm. La séparatrice est constituée par une lame d'air d'épaisseur réglable, comprise entre deux prismes de téflon dont les faces sont sous incidence brewstérienne. L'échantillon solide ou liquide forme une lame à faces parallèles placée contre le fond d'une cuve métallique, à l'extrémité de l'un des bras de l'interféromètre. La précision obtenue est fonction inverse de l'absorption diélectrique dans l'échantillon ; pour des pertes moyennes l'erreur reste inférieure à 3 % sur ε' et 6 % sur ε". Dans le cas du chlorure de tertio-butyle congelé à - 35 °C on a trouvé ε' = 2,7 et ε" = 1,7.

PACS
7310K - Dielectric variables measurement.

Key words
dielectric loss measurement -- instrumentation -- permittivity measurement